Home > A- Z Directory-EW Sector > Air Quality Monitoring

Return to top
Return to top
Return to top
Return to top
Return to top
 

For additional information contact:

1-800-264-7477

michael.vanwart@edocorp.com

 

 

Air Quality Monitoring
The RAM 2000™ is an affordable, user friendly instrument that can provide real-time identification and quantification of over 250 chemicals and detection levels mostly in the range of 0.1 to 15 ppb (path-averaged). Unlike point monitors, it monitors an entire path up to 1000 meters in length, giving a vapor cloud little chance to pass through undetected.

The RAM 2000™ is the most advanced, practical method for gas monitoring along a fenceline. This Open-Path FTIR Analyzer can operate continuously and unattended while identifying and quantifying individual chemicals in gas mixtures.

When gases pass through the monitored path, the RAM 2000™ detects, quantifies, and reports the amounts of each chemical in real-time.

RAM 2000™ Applications

RAM 2000™ Advantages

OPTIONS

* The use of the ETV® name or logo does not imply approval or certification of the product by the EPA nor does it make any claims, explicit or implied,
as to product performance. The ETV® report is available at
http://www.epa.gov/etv/pdfs/vrvs/01_vr_ail.pdf

SYSTEM SOFTWARE

The RAM 2000™ unit is a remote chemical detection system used by petrochemical and pharmaceutical companies to monitor chemical emissions. The RAM 2000™ software, RMMsoft™, controls the RAM 2000™ system. State-of-the-practice software tools and methods, a SEI-compliant development process combined with advanced signal processing algorithms and an easy to use GUI interface resulted in the development of this easy to operate system.

The RMMsoft™ software was awarded the Long Island Software Award (LISA) for the most innovative software from the Long Island Software and Technology Network (LISTnet). We won for the large company category. LISTnet is an organization comprised of over 330 companies dedicated to promoting the Long Island Software industry.

Windows-based RMMsoft™ facilitates EASY-TO-USE OPERATION by nontechnical personnel as well as detailed spectroscopic analysis.

SPECIFICATIONS

CONFIGURATION   Monostatic FTIR spectrometer for fenceline, open-path and remediation applications
INTERFEROMETER   Continuous scan Michaelson interferometer
Dual porch swing mirror suspension
Manually-adjustable fixed mirrors
HeNe laser frequency calibration
RANGE   Up to 500 m distance, spectrometer to retroreflector
SPECTRAL RANGE   700 to 4,000 cm-1
RESOLUTION   0.5 cm-1 maximum
MINIMUM DETECT
LEVELS (MDL)
  Detection limits for chemicals measured by open-path FTIR systems will vary depending on the chemical, atmospheric conditions (humidity and temperature) and whether interfering chemicals may be present. The range of detection limits for a 100 meter separation between the transmitter/receiver telescope and the retroreflector is from 0.1 to 15 ppb for most infrared-active chemicals.
SCAN TIME   1.7 seconds per scan nominal at 0.5 cm-1 resolution
NOISE   Measured as noise equivalent absorbance in milliabsorbance units (mA.U.) for 64 co-adds at 100 m range
968-1008 cm-1 0.10 mA.U.
2480-2520 cm-1 0.01 mA.U.
4380-4420 cm-1 0.40 mA.U.
SOFTWARE   Interactive, user friendly, Windows-based, knowledge of spectroscopy is not required
WEATHERIZED   Temperature controlled interferometer assembly, infrared source and MCT detector for increased stability
Enclosure gasket sealed from weather
Inlet valve port provided for dry gas purging and pressurization
Desiccant cartridge provided to remove humidity and maintain enclosure dryness, removable from outside for easy replacement
Nonhygroscopic ZnSe used for all optical windows and beamsplitter
DETECTOR   Mercury cadmium telluride (MCT) detector element
Liquid nitrogen cooled for minimal noise contribution
Vacuum dewar with six to ten hour refill cycle
LN2 automatic dewar refill system, 14 to 28 day unattended operation, depending on supply dewar size or direct connection to a LN2 supply line
QA CELL   Integrated 15-cm QA cell with quick disconnect fittings and a safety pressure relief valve provided. ZnSe cell windows are replaceable
TELESCOPE   Newtonian with a 10-inch aperture
RETROREFLECTOR   Corner cubes reduce system alignment sensitivity
State-of-the-art durable mirror surface coating, not gold
Accuracy of cubes: 5 arcseconds
Number of cubes: 37
Size of array: 17-inch diameter
Nema case with cover
Tripod mount adapter
Option 1 - Portable tripod with manual pan/tilt mount
Option 2 - Fixed mount wall and/or pole

Privacy Policy | Copyright© 2007 | Webmaster

 

 

System Software - RMMsoft™
Specifications
   

Related Topic:
Liquid Screening Portal - RAM2000™

Liquid Screening Portal - RAM2000™
Brochure Download
   